Iron Analyzer Liquiline System CA80FE

Brand: Endress+Hauser

Iron Analyzer Liquiline System CA80FE

Brand: Endress+Hauser

Local Support: Authorized Distributor Serving Alabama, Mississippi, Louisiana, Florida, New York, Connecticut, Massachusetts, Vermont, New Hampshire, Maine, Rhode Island, Puerto Rico, US Virgin Islands and parts of Tennessee and Arkansas.

Extended Extended

Liquiline System CA80FE offers precise, regulation-compliant online measurement. It supports you in controlling the iron content of drinking water, mineral water or source water for industrial processes. Like all Liquiline System analyzers, it enables plug & play of up to four Memosens sensors – minimizing the installation effort. Automatic calibration and cleaning and the low reagent consumption reduce operating costs while advanced diagnostics with remote access ensure process safety.

Liquiline System CA80FE monitors:

  • Iron content in drinking water and mineral water
  • Residual iron after precipitation, flocculation and filtration processes
  • Process water containing diluted iron
Features
  • Compliance with standard ferrozine method ensures direct comparability to lab measurements
  • Supports predictive maintenance by advanced diagnostics
  • Reduced operating costs through automatic calibration and cleaning as well as low reagent consumption
  • Easy upgrade of functionality - even to a complete measuring station - simply by adding modules and connecting Memosens sensors.
  • Comfortable remote access via integrated web server
  • Seamless integration into process control systems thanks to digital field busses such as Modbus, PROFIBUS or EtherNet/IP

Specifications

Product Family

  • Liquiline

Measuring Principle

  • Colorimetric

Characteristic

  • Process analyzer for iron in aqueous solutions
  • Comply with standard colorimetric measuring principle, ferrozine method

Size

  • Housing (open version): 793 x 530 x 417 mm (31.22 x 20.87 x 16.42 in)
  • Housing (closed version): 793 x 530 x 463 mm (31.22 x 20.87 x 18.23 in)
  • Housing with base: 1723 x 530 x 463 mm (67.83 x 20.87 x 18.23 in)

Design

  • Open design, cabinet and stand housing
  • High-Performance plastic ASA-PC, additional stand coated steel

Process Temperature

  • 4 to 40 °C (39.2 to 104 °F)

Ambient Temperature

  • 5 to 40 °C (41 to 104 °F)
  • Outdoor version: - 20 to 40 °C (-4 to 104 °F)

Process Pressure

  • At atmospheric pressure, < 0.2 bar

Sample Flow Rate

  • Min. 5 ml/min (0.17 fl.oz/min)

Consistency of the Sample

  • Suspended solids content
  • Turbitity < 50 NTU, aqueous, homogenized

Application

  • Monitoring of the iron content of potable water, mineral water and source water for industrial processes

Power Supply

  • 100 to 120 VAC / 200 to 240 VAC ± 10%
  • 24 VDC ± 10%
  • 50 ± 1 or 60 ± 1.2 Hz

Output / Communication

  • 2x 0/4 to 20 mA
  • Webserver, Modbus, Ethernet/IP, Profibus DP

Input

  • 1 or 2 measuring channel
  • 1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)

Measuring Range

  • 0.05 to 2 mg/l Fe
  • 0.1 to 5 mg/l Fe
  • 0.1 to 5 mg/l with dilution function to maximum 2.0 to 100 mg/l Fe

Consumables

  • Reagents and standard solutions CY80FE for the operation
  • Regular maintenance is done with the parts of the maintenance kit CAV800
show more