Iron Analyzer Liquiline System CA80FE

Brand: Endress+Hauser

Iron Analyzer Liquiline System CA80FE

Brand: Endress+Hauser
Extended Extended

Liquiline System CA80FE offers precise, regulation-compliant online measurement. It supports you in controlling the iron content of drinking water, mineral water or source water for industrial processes. Like all Liquiline System analyzers, it enables plug & play of up to four Memosens sensors – minimizing the installation effort. Automatic calibration and cleaning and the low reagent consumption reduce operating costs while advanced diagnostics with remote access ensure process safety.

Liquiline System CA80FE monitors:

  • Iron content in drinking water and mineral water
  • Residual iron after precipitation, flocculation and filtration processes
  • Process water containing diluted iron
Features
  • Compliance with standard ferrozine method ensures direct comparability to lab measurements
  • Supports predictive maintenance by advanced diagnostics
  • Reduced operating costs through automatic calibration and cleaning as well as low reagent consumption
  • Easy upgrade of functionality - even to a complete measuring station - simply by adding modules and connecting Memosens sensors.
  • Comfortable remote access via integrated web server
  • Seamless integration into process control systems thanks to digital field busses such as Modbus, PROFIBUS or EtherNet/IP

Specifications

Measuring Principle

  • Colorimetric

Characteristic

  • Process analyzer for iron in aqueous solutions
  • Comply with standard colorimetric measuring principle, ferrozine method

Size

  • Housing (open version): 793 x 530 x 417 mm (31.22 x 20.87 x 16.42 in)
  • Housing (closed version): 793 x 530 x 463 mm (31.22 x 20.87 x 18.23 in)
  • Housing with base: 1723 x 530 x 463 mm (67.83 x 20.87 x 18.23 in)

Design

  • Open design, cabinet and stand housing
  • High-Performance plastic ASA-PC, additional stand coated steel

Process Temperature

  • 4 to 40 °C (39.2 to 104 °F)

Ambient Temperature

  • 5 to 40 °C (41 to 104 °F)
  • Outdoor version: - 20 to 40 °C (-4 to 104 °F)

Process Pressure

  • At atmospheric pressure, < 0.2 bar

Sample Flow Rate

  • Min. 5 ml/min (0.17 fl.oz/min)

Consistency of the Sample

  • Suspended solids content
  • Turbitity < 50 NTU, aqueous, homogenized

Application

  • Monitoring of the iron content of potable water, mineral water and source water for industrial processes

Power Supply

  • 100 to 120 VAC / 200 to 240 VAC ± 10%
  • 24 VDC ± 10%
  • 50 ± 1 or 60 ± 1.2 Hz

Output / Communication

  • 2x 0/4 to 20 mA
  • Webserver, Modbus, Ethernet/IP, Profibus DP

Input

  • 1 or 2 measuring channel
  • 1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)

Measuring Range

  • 0.05 to 2 mg/l Fe
  • 0.1 to 5 mg/l Fe
  • 0.1 to 5 mg/l with dilution function to maximum 2.0 to 100 mg/l Fe

Consumables

  • Reagents and standard solutions CY80FE for the operation
  • Regular maintenance is done with the parts of the maintenance kit CAV800
show more