Digital Conductivity Sensor Memosens CLS15E
Brand: Endress+HauserDigital Conductivity Sensor Memosens CLS15E
Brand: Endress+Hauser
Memosens CLS15E is perfectly suited for conductivity measurement in applications with low measuring ranges such as boiler feedwater and chip cleaning. The contacting conductivity sensor performs reliably and accurately even in hazardous areas. CLS15E features Memosens 2.0 digital technology, offering extended storage of calibration, adjustment and process data. It facilitates predictive maintenance and provides the perfect basis for IIoT services.
Memosens CLS15E measures conductivity in pure and ultrapure water applications for the monitoring and control of:
- Boiler feedwater
- Ion exchangers
- Reverse osmosis
- Cooling water
- Distillation
- Chip cleaning
The digital sensor can be used in power plants for:
- Degassed conductivity
- pH measurement with differential conductivity
With its international explosion protection approvals, such as ATEX and FM, the sensor is suitable for hazardous area use.
Features- Designed for low maintenance and a long operating life, the sensor offers best value for money.
- The sensor is sterilizable and autoclavable, allowing for installation in sterile plants.
- Thanks to its electrode geometry, Memosens CLS15E provides reliable and accurate measured values at low conductivities.
- A quality certificate stating the individual cell constant enables perfect adjustment of the measuring point.
- IIoT ready: Memosens 2.0 offers extended storage of calibration and process data, enabling better trend identification and providing a future-proof basis for predictive maintenance and enhanced IIoT services.
- Non-contact inductive signal transmission ensures maximum process safety.
Specifications
Brand
Measuring Principle
- Conductive
Application
- Measurement in pure and ultrapure water range
- Monitoring of ion exchangers
- Reverse osmosis
- Distillation and chip cleaning
Characteristic
- Digital 2-electrode conductivity sensor
Measuring Range
- k = 0.01: 0.04 to 20 µS/cm
- k = 0.1: 0.10 to 200 µS/cm
Design
- 2-electrode conductivity cell with coaxially arranged electrodes, electropolished
Material
- Electrode: stainless steel 1.4435
- Electrode shaft: PES
Dimension
- Electrode diameter: 16 mm (0.63 inch)
- Electrode length: appr. 55 mm (2.17 inch)
Process Temperature
- Threaded with fixed cable: -20 to 100 °C (-4 to 212 °F)
- Threaded with plug-in head: -20 to 120 °C (-4 to 248 °F)
- Sterilization: max. 140 °C (284 °F) for 30 minutes
Process Pressure
- 13 bar at 20 °C (188 psi at 68 °F) absolute
- 2 bar at 120 °C (14 psi at 248 °F) absolute
Temperature Sensor
- Pt1000
Ex Certification
- ATEX, NEPSI, CSA, IECEx, INMETRO, EAC Ex
Connection
- Process: 1/2" and 3/4" NPT, Clamp 1-1/2"
- Sensor connection: Inductive, digital connection head with Memosens 2.0 technology
Ingress Protection
- IP68
Additional Certifications
- Calibration certification of the cell constante and temperature, material certification 3.1