Silica Analyzer Liquiline System CA80SI
Brand: Endress+HauserSilica Analyzer Liquiline System CA80SI
Brand: Endress+Hauser
Liquiline System CA80SI provides precise online analysis of silica and helps protect plant equipment from glass-like deposits ensuring optimum performance of turbines and ion exchangers. Its unique combination of peristaltic and high-precision dispenser pumps enables extremely reliable operation and low maintenance. The silica analyzer offers advanced diagnostics and can be easily adapted to your process needs by retrofitting up to 6 channels and connecting up to 4 Memosens sensors.
The CA80SI silica analyzer monitors trace levels of silica in the power industry and utilities:
- Indication of the purity of boiler feed water and steam
- Provision of early warning for condenser leaks
- Early warning of exhaustion of ion exchangers
- Protection of expensive plant equipment in the power industry: The silica analyzer monitors trace levels of silica in water steam cycles helping to avoid glass-like deposits on the equipment surfaces
- Early online detection of critical silica levels ensures efficient power plant performance and optimizes maintenance and blowdown cycles by early counter measures
- Standard heteropoly blue method allows direct comparability to lab results
- Best reliability: Unique combination of peristaltic and high-precisions dispenser pumps ensures stable operation and low maintenance
- Optimized investment: The option of up to 6 sample channels meets changing process needs at any time, and connection of Memosens sensors allows an easy upgrade to a complete measuring station
- Seamless integration into process control systems thanks to digital field buses such as Modbus, EtherNet/IP or PROFIBUS
- Advanced diagnostics and remote access via web server enable fast remedy in case of errors
Specifications
Brand
Measuring Principle
- Colorimetric
Characteristic
- Analyzer for silica in aqueous solutions
Measuring Method
- Comply with standard colorimetric measuring principle - heteropoly blue method
Size
- Housing (open version): 793 x 530 x 417 mm (31.22 x 20.87 x 16.42 in)
- Housing (closed version): 793 x 530 x 463 mm (31.22 x 20.87 x 18.23 in)
- Housing with base: 1723 x 530 x 463 mm (67.83 x 20.87 x 18.23 in)
Design
- Open design / cabinet
- Plastic ASA-PC
Process Temperature
- 5 to 45 °C (41 to 113 °F)
Ambient Temperature
- 5 to 40 °C (41 to 104 °F)
Process Pressure
- 1 to 5 bar (14.5 to 72.5 psi)
Sample Flow Rate
- 60 to 250 ml/min (2.03 to 8.45 fl.oz/min)
Consistency of the Sample
- Particle free
Specials
- Easy upgrade to measuring station with up to four digital Memosens sensors
- Automatic calibration and cleaning
- User-configurable measuring, cleaning and calibration intervals
- Optional up to 6 sample channels
- Flexible functionality and modular expandable
- Digital communication for remote access
Application
- Determination of silica concentration in boiler feed water
- Monitoring of water-steam-circuit and condensate
- Monitoring and optimization of reverse osmosis and desalination
Power Supply
- 100 to 120 VAC / 200 to 240 VAC ± 10%
- 50 ± 1 or 60 ± 1.2 Hz
Output / Communication
- 2x 0/4 to 20 mA
- Optional: Webserver, Modbus, Ethernet/IP, Profibus DP
Input
- 1, 2, 4 or 6 measuring channels
- Optional 1 to 4 digital inputs for sensors with Memosens protocol
Measuring Range
- 0 to 500 µg/l (ppb)
- 50 to 5000 µg/l (ppb)
Consumables
- Reagents and standard solutions CY80SI
- Maintenance kit CAV800