TA572 Barstock Thermowell

Brand: Endress+Hauser

TA572 Barstock Thermowell

Brand: Endress+Hauser
Extended Extended

Due to the challenging process conditions by heavy duty applications the load capacity of a thermowell must be calculated exactly. Dye penetration tests, ultrasound test, helium leakage test, pressure endurance test as well as various, non-destructive material tests prove the quality of materials and processing.

TA572 is designed for use with resistance and thermocouple thermometers mainly in heavy duty applications.

Features
  • The extension, the immersion and the tapering lengths as well as the bar dimensions can be chosen according to process requirements
  • A wide choice of standard materials is available; additional versions can be ordered according to specification
  • Different grades of surface finishing are also available
  • The process connection is welded. The thermowell stem shape is conical

Specifications

Measuring Principle

  • Bar stock Thermowell

Characteristic / Application

  • metric style
  • weld-in process connections
  • round extension

Head Connection

  • internal thread: 1/2" NPT, 1/2" NPSC

Maximum Standard Immersion Length

  • 900 mm (35.43")

Max. Immersion Length on Request

  • 5,000 mm (196.85")

Process Connection

  • to weld-in

Thermowell Root Diameter

  • 22 mm (0.87")
  • 27 mm (1.06")

Medium Contact Material

  • 1.4401 (316)
  • 1.4404 (316L)
  • 1.4571 (316Ti)

Wetted Part Finishing (Ra)

  • < 0.8 μm (31.50 μin)
  • < 1.6 μm (63.00 μin)

Tip Shape

  • straight
  • conical
  • conical tapered

Temperature Range

  • -200...700 °C (-328...1.292 °F)

Max. Process Pressure (static)

  • 500 bar (7252 psi)

Max. Process Pressure at 400 °C

  • 300 bar (4351 psi)
show more