Proline Promass H 100 Coriolis Flowmeter

Brand: Endress+Hauser

Proline Promass H 100 Coriolis Flowmeter

Brand: Endress+Hauser
Xpert Xpert
Heartbeat Technology Heartbeat Technology

The chemically resistant single-tube design of the Promass H is destined for applications requiring highest corrosion resistance. Combined with the smallest transmitter housing available today it delivers full performance on the smallest footprint. Designed for applications where space is a premium, Promass H 100 will be the preferred choice for system integrators, skid builders and equipment manufacturers.

  • Measuring principle operates independently of physical fluid properties such as viscosity or density
  • Highly accurate measurement of liquids and gases in applications requiring highest corrosion resistance

Device properties

  • Measuring tube made of Tantalum, Zirconium
  • Nominal diameter: DN 8 to 50 (3/8 to 2")
  • Medium temperature up to +205 °C (+401 °F)
  • Robust, ultra-compact transmitter housing
  • High ingress protection: IP69K
  • Local display available
Features
  • Maximum safety for chemically aggressive fluids – corrosion-resistant wetted parts
  • Fewer process measuring points – multivariable measurement (flow, density, temperature)
  • Space-saving installation – no in/outlet run needs
  • Space-saving transmitter – full functionality on the smallest footprint
  • Time-saving local operation without additional software and hardware – integrated web server
  • Integrated verification – Heartbeat Technology

Specifications

Measuring Principle

  • Coriolis

Product Headline

  • Chemically resistant single-tube flowmeter with an ultra-compact transmitter
  • Measuring highly accurately liquids and gases in applications requiring highest corrosion resistance

Sensor Features

  • Maximum safety for chemically aggressive fluids – corrosion‐resistant wetted parts
  • Fewer process measuring points – multivariable measurement (flow, density, temperature)
  • Space‐saving installation – no in-/outlet run needs
  • Measuring tube made of Tantalum, Zirconium. Nominal diameter: DN 8 to 50 (3/8 to 2"). Medium temperature up to +205 °C (+401 °F)

Transmitter Features

  • Space-saving transmitter – full functionality on the smallest footprint
  • Time‐saving local operation without additional software and hardware – integrated web server
  • Integrated verification – Heartbeat Technology
  • Robust, ultra-compact transmitter housing. Highest degree of protection: IP69. Local display available

Nominal Diameter Range

  • DN 8 to 50 (3/8 to 2")

Wetted Materials

  • Measuring tube: Tantalum 2.5W; 702 (UNS R60702)
  • Connection: Tantalum; 702 (UNS R60702)

Measured Variables

  • Mass flow, density, temperature, volume flow, corrected volume flow, reference density, concentration

Max. Measurement Error

  • Mass flow (liquid): ±0.1 %
  • Volume flow (liquid): ±0.1 %
  • Mass flow (gas, Tantalum only): ±0.5 %
  • Density (liquid): ±0.0005 g/cm³

Measuring Range

  • 0 to 70,000 kg/h (0 to 2570 lb/min)

Max. Process Pressure

  • PN 40, Class 300, 20K

Medium Temperature Range

  • Tantalum: –50 to +150 °C (–58 to +302 °F)
  • Zirconium: –50 to +205 °C (–58 to +401 °F)

Ambient Temperature

  • Standard: –40 to +60 °C (–40 to +140 °F)
  • Option: –50 to +60 °C (–58 to +140 °F)

Sensor Housing Material

  • 1.4301 (304), corrosion resistant

Transmitter Housing Material

  • Compact: AlSi10Mg, coated
  • Compact/ultra‐compact: 1.4301 (304)

Degree of Protection

  • Standard: IP66/67, type 4X enclosure
  • Option: IP69

Display/Operation

  • 4‐line backlit display available (no local operation)
  • Configuration via web browser and operating tools possible

Outputs

  • 4‐20 mA HART (active)
  • Pulse/frequency/switch output (passive)

Inputs

  • None

Digital Communication

  • HART, Modbus RS485, EtherNet/IP, PROFIBUS DP, PROFINET

Power Supply

  • DC 20 to 30 V

Hazardous Area Approvals

  • ATEX, IECEx, cCSAus, INMETRO, NEPSI, EAC

Product Safety

  • CE, C-Tick, EAC marking

Metrological Approvals and Certificates

  • Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
  • Heartbeat Technology complies with the requirements for traceable verification cording to ISO 9001:2008 – Section 7.6a (TÜV SÜD attestation)

Pressure Approvals and Certificates

  • PED, CRN

Material Certificates

  • 3.1 material
show more