Heartbeat Technology Heartbeat Technology

The highly accurate Promass H is destined for applications requiring maximum corrosion resistance and guarantees optimal safety for chemically aggressive fluids. With its compact transmitter Promass H 300 offers high flexibility in terms of operation and system integration: access from one side, remote display and improved connectivity options. Heartbeat Technology enables process safety at all times.

  • Measuring principle operates independently of physical fluid properties such as viscosity or density
  • Highly accurate measurement of liquids and gases in applications requiring highest corrosion resistance

Device properties

  • Measuring tube made of Tantalum, Zirconium
  • Nominal diameter: DN 8 to 50 (3/8 to 2")
  • Medium temperature up to +205 °C (+401 °F)
  • Compact dual-compartment housing with up to 3 I/Os
  • Backlit display with touch control and WLAN access
  • Remote display available
Features
  • Maximum safety for chemically aggressive fluids – corrosion-resistant wetted parts
  • Fewer process measuring points – multivariable measurement (flow, density, temperature)
  • Space-saving installation – no in-/outlet run needs
  • Full access to process and diagnostic information – numerous, freely combinable I/Os and Ethernet
  • Reduced complexity and variety – freely configurable I/O functionality
  • Integrated verification – Heartbeat Technology

Specifications

Measuring Principle

  • Coriolis

Product Headline

  • Chemically resistant single-tube flowmeter with a compact, easily accessible transmitter
  • Highly accurate measurement of liquids and gases in applications requiring highest corrosion resistance

Sensor Features

  • Maximum safety for chemically aggressive fluids – corrosion‐resistant wetted parts
  • Fewer process measuring points – multivariable measurement (flow, density, temperature)
  • Space‐saving installation – no in-/outlet run needs
  • Measuring tube made of Tantalum, Zirconium. Nominal diameter: DN 8 to 50 (3/8 to 2"). Medium temperature up to +205 °C (+401 °F)

Transmitter Features

  • Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses
  • Reduced complexity and variety – freely configurable I/O functionality
  • Integrated verification – Heartbeat Technology
  • Compact dual-compartment housing with up to 3 I/Os. Backlit display with touch control and WLAN access. Remote display available

Nominal Diameter Range

  • DN 8 to 50 (3/8 to 2")

Wetted Materials

  • Measuring tube: Tantalum 2.5W; 702 (UNS R60702)
  • Connection: Tantalum; 702 (UNS R60702)

Measured Variables

  • Mass flow, density, temperature, volume flow, corrected volume flow, reference density, concentration

Max. Measurement Error

  • Mass flow (liquid): ±0.10 %
  • Volume flow (liquid): ±0.10 %
  • Mass flow (gas, Tantalum only): ±0.50 %
  • Density (liquid): ±0.0005 g/cm³

Measuring Range

  • 0 to 70,000 kg/h (0 to 2,570 lb/min)

Max. Process Pressure

  • PN 40, Class 300, 20K

Medium Temperature Range

  • Tantalum: –50 to +150 °C (–58 to +302 °F)
  • Zirconium: –50 to +205 °C (–58 to +401 °F)

Ambient Temperature

  • Standard: –40 to +60 °C (–40 to +140 °F)
  • Option: –50 to +60 °C (–58 to +140 °F)

Sensor Housing Material

  • 1.4301 (304), corrosion resistant

Transmitter Housing Material

  • AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L

Degree of Protection

  • IP66/67, type 4X enclosure

Display/Operation

  • 4-line backlit display with touch control (operation from outside)
  • Configuration via local display and operating tools possible
  • Remote display available

Outputs

  • 3 outputs:
  • 4-20 mA HART (active/passive)
  • 4-20 mA WirelessHART
  • 4-20 mA (active/passive)
  • Pulse/frequency/switch output (active/passive)
  • Double pulse output (active/passive)
  • Relay output

Inputs

  • Status input
  • 4-20 mA input

Digital Communication

  • HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA

Power Supply

  • DC 24 V
  • AC 100 to 230 V
  • AC 100 to 230 V / DC 24 V (non-hazardous area)

Hazardous Area Approvals

  • ATEX, IECEx, cCSAus, NEPSI, INMETRO, EAC, UK Ex, KC

Product Safety

  • CE, C-tick, EAC marking

Functional Safety

  • Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511

Metrological Approvals and Certificates

  • Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
  • Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)

Pressure Approvals and Certificates

  • PED, CRN

Material Certificates

  • 3.1 material
show more