iTHERM FlameLine TAF16 High Temperature Thermometer

Brand: Endress+Hauser

iTHERM FlameLine TAF16 High Temperature Thermometer

Brand: Endress+Hauser
Extended Extended

Measuring points with extremely high process temperatures require this high-temperature thermometer with a ceramic or metal thermowell as well as thermocouples made from special metals such as platinum and rhodium. The thermowell made from special materials protects from mechanical and chemical damages in the process and increases the life span of the used sensors. This leads to: Cost savings for maintenance of the measuring point, quality improvements of the products and increased plant safety.

  • Applicable for temperature measurement in cement production, steel treatment, incinerators and fluidized bed furnaces. The TAF16 contains a single or double thermocouple insert and a metal or ceramic protection tube as well as an optional internal sheath
  • Process temperatures: up to +1,700 °C (+3,092 °F)
Features
  • Long lifetime by usage of innovative thermowell materials with increased wear and chemical resistance
  • Long-term stable measurement due to sensor protection with non-porous materials
  • Flexible product selection thanks to modular design
  • Optimized life cycle costs by means of replaceable spare parts

Specifications

Measuring Principle

  • Thermocouple

Characteristic / Application

  • metric style
  • modular temperature assembly
  • high temperature probe (furnace)
  • incl. thermowell / protection tube (metal)

Thermowell / Protection Tube

  • welded protection tube
  • ceramic protection tube

Insert / Probe

  • mineral insulated (MI), flexible
  • ceramic capilarry, rigid

Outer Diameter Protection Tube / Insert

  • 14.0 mm (0.55")
  • 15.0 mm (0.59")
  • 17.2 mm (0.68")
  • 18.0 mm (0.71")
  • 21.3 mm (0.84")
  • 22.0 mm (0.87")
  • 26.7 mm (1.05")
  • 28.0 mm (1.10")

Max. Immersion Length on Request

  • up to 4,525.00 mm (178.15'')

Material Protection Tube/ Thermowell

  • 1.4401 (316)
  • 1.4749 (~446)
  • 1.4841 (310)
  • Alloy 600 (2.4816)
  • Alloy 601 (2.4851)
  • Incoloy 800HT (1.4959)
  • Kanthal AF
  • Kanthal Super
  • NiCo
  • Siliciumnitrid(SiN)

Process Connection

  • compression fitting: G1/2", G3/4", G1", G1-1/4", G1-1/2"
  • adjustable Flange

Tip Shape

  • straight

Operating Temperature

  • Type K: -40 °C ...1,100 °C (-40 °F ...2,012 °F)
  • Type J: -40 °C ...750 °C (-40 °F ...1,382 °F)
  • Type N: -40 °C ...1,150 °C (-40 °F ...2,102 °F)
  • Type S: 0 °C ...1,400 °C (32 °F ...2,552 °F)

Max. Process Pressure (static)

  • at 20 °C: 1 bar (15 psi)

Accuracy

  • class 2 acc. to IEC 60584

Integration Head Transmitter

  • yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)

Certification

  • SIL (transmitter only)
show more