Aluminum Analyzer Liquiline System CA80AL
Brand: Endress+HauserAluminum Analyzer Liquiline System CA80AL
Brand: Endress+Hauser
Liquiline System CA80AL offers precise, regulation-compliant online measurement of aluminum residues in drinking water and wastewater.
The analyzer helps you to optimize aluminum dosing in flocculation processes and phosphate removal. It enables plug & play of up to 4 Memosens sensors – minimizing the installation effort.
Automatic calibration and cleaning and the low reagent consumption reduce operating costs while advanced diagnostics with remote access ensure process safety.
Liquiline System CA80AL aluminum analyzer supports:
- Limit monitoring in drinking water
- Optimization of aluminum-based coagulation processes
- Phosphate elimination in wastewater treatment plants
- The aluminum analyzer uses the pyrocatechol violet method according to DIN ISO 10566, ensuring direct comparability to lab results
- Supports predictive maintenance by advanced diagnostics
- Reduces operating costs thanks to automatic calibration and cleaning as well as low reagent consumption
- Easy upgrade of functionality to a complete measuring station - by adding modules and connecting Memosens sensors.
- Comfortable remote access via integrated web server
- Seamless integration into process control systems thanks to digital field buses such as Modbus, PROFIBUS, EtherNet/IP
Specifications
Brand
Measuring Principle
- Colorimetric
Characteristic
- Process analyzer for aluminum in aqueous solutions
- Comply with standard colorimetric measuring principle - pyrocatechol violet methot
Size
- Housing (closed version): 793 x 530 x 463 mm (31.22 x 20.87 x 18.23 in)
- Housing with base: 1723 x 530 x 463 mm (67.83 x 20.87 x 18.23 in)
- Housing (open version): 793 x 530 x 417 mm (31.22 x 20.87 x 16.42 in)
Design
- Open design, cabinet and stand housing
- High-Performance plastic ASA-PC, additional stand coated steel
Process Temperature
- 4 ... 40 °C (39.2 ... 104 °F)
Ambient Temperature
- 5 ... 40 °C ( 41 ... 104 °F)
- Outdoor version: - 20 to 40 °C (-4 to 104 °F)
Process Pressure
- at atmospheric pressure, < 0.2 bar
Sample Flow Rate
- min. 5ml/min (0.17 fl.oz/min)
Consistency of the Sample
- low solid content (< 50NTU), aqueous homogeneous
Application
- Monitoring of the aluminum content of potable water, mineral water and industrial waste water
Power Supply
- 100 ... 120 VAC / 200 ... 240 VAC ± 10%
- 24 VDC ± 10%
- 50 ± 1 oder 60 ± 1.2 Hz
Output / Communication
- 2x 0/4 ... 20 mA
- Webserver, Modbus, Ethernet/IP, Profibus DP
Input
- 1 or 2 measuring channel
- 1 ... 4 digital sensor inputs for sensors with Memosens protocol (optional)
Measuring Range
- 15 to 1,000 µg/l Al
- 15 to 1,000 µg/l with dilution function to maximum 300 to 20,000 µg/l Al
Consumables
- Reagents and standard solutions CY80AL as well as cleaner CY800 are necessary for the operation. Regular maintenance is done with the parts of the maintenance kit CAV800