Proline T-Mass B 150 Thermal Mass Flowmeter

Brand: Endress+Hauser

Proline T-Mass B 150 Thermal Mass Flowmeter

Brand: Endress+Hauser
Lean Lean

The t-mass B 150 insertion version is suitable for large pipelines or rectangular ventilation ducts. It is designed for the cost-effective measurement of utility gases, in particular compressed air. It is a trending device aimed at sub-metering applications. Its 4-wire technology is contained within a rugged compact aluminum housing. Customer-specific settings are saved on the display and can be transferred from one device to another by means of the display.

  • Measuring principle is characterized by a high operable flow range and direct mass flow measurement
  • Suitable for air, nitrogen, carbon dioxide and argon in circular piping or rectangular ducts

Device properties

  • Insertion version for nominal diameter DN 80 to 1500 (3 to 60")
  • Installation and removal of sensor without process interruption
  • Easy installation
  • Device in compact version with DC 24 V power supply
  • 4-20 mA HART, pulse/frequency/switch output
  • Compact and robust transmitter
Features
  • Suitable for air, nitrogen, carbon dioxide and argon in circular piping or rectangular ducts
  • Optimal process monitoring – easy measurement even at low pressures and flow velocities
  • Cost-effective measurement – easy installation, negligible pressure loss and maintenance-free
  • Reliable flow trending – multivariable measurement
  • Fast and efficient commissioning – guided operating menus
  • High plant availability – self-diagnostics and error monitoring
  • Automatic recovery of data for servicing

Specifications

Measuring Principle

  • Thermal

Product Headline

  • The flowmeter for cost-effective measurement and easy monitoring of utility gases
  • Suitable for air, nitrogen, carbon dioxide and argon in circular piping or rectangular ducts

Sensor Features

  • Optimal process monitoring – easy measurement even at low pressures and flow velocities. Cost‐effective measurement – easy installation, negligible pressure loss and maintenance‐free. Reliable flow trending – multivariable measurement
  • Insertion version for nominal diameter DN 80 to 1500 (3 to 60"). Installation and removal of sensor without process interruption. Easy installation

Transmitter Features

  • Fast and efficient commissioning – guided operating menus. High plant availability – self‐diagnostics and error monitoring. Automatic recovery of data for servicing
  • Device in compact version with DC 24 V power supply. 4-20 mA HART, pulse/frequency/switch output. Compact and robust transmitter

Nominal Diameter Range

  • DN 80 to 1500 (3 to 60")

Wetted Materials

  • Transducer: 1.4404 (316L)
  • Insertion tube: 1.4404 (316L); 1.4435 (316L)
  • Compression fitting: 1.4404 (316L)
  • Sealing ring: EPDM; HNBR; 1.4401 (316)
  • Clamping ring: PEEK 450G

Measured Variables

  • Mass flow, temperature, corrected volume flow, FAD volume flow

Max. Measurement Error

  • 3 % o.r.
  • 4 % o.r.
  • 5 % o.f.s.
  • depending on chosen option of ordering feature "Calibration flow"

Measuring Range

  • 20 to 720,000 kg/h (45 to 1,587,600 lb/h)
  • 20 to 1,080,000 kg/h (45 to 2,381,400 lb/h)
  • for air, depending on chosen option of ordering feature "Calibration flow"

Max. Process Pressure

  • 20 bar g (290 psi g)

Medium Temperature Range

  • –40 to +100 °C (–40 to +212 °F)

Ambient Temperature

  • –40 to +60 °C (–40 to +140 °F)

Transmitter Housing Material

  • AlSi10Mg, coated

Degree of Protection

  • IP66/67, type 4X enclosure

Display/Operation

  • 4‐line display with push Buttons
  • Configuration via local display and operating tools possible

Outputs

  • 4‐20 mA HART (active)
  • Pulse/frequency/switch output (passive)

Inputs

  • Status input

Digital Communication

  • HART

Power Supply

  • DC 18 to 30 V

Hazardous Area Approvals

  • ATEX, IECEx, cCSAus

Metrological Approvals and Certificates

  • Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR

Pressure Approvals and Certificates

  • CRN
show more