Raman Calibration Tool

Brand: Endress+Hauser

Raman Calibration Tool

Brand: Endress+Hauser
Extended Extended

The Raman Calibration Tool (RCT) offers laboratory quality calibration in a field-capable tool, providing high accuracy critical for quantitative analysis. Coupled with wavelength (x-axis) and laser calibrations that are integrated into Raman analyzers, the Raman Calibration Tool provides the highest level of accuracy for even the most demanding Raman measurements. Used routinely, the RCT protocol facilitates calibration transfer between laboratory instruments and on-line process analyzers.

  • LNG truck loading
  • LNG rundown to storage tanks
Features
  • Robust calibration protocol
  • Excellent system accuracy and unparalleled instrument precision
  • Minimal technical support required
  • Incorporates NIST Standard Reference Material (SRM); Compatible with ASTM E2911
  • Facilitates easy calibration transfer
  • Chemical methods and models instrument-independent
  • Simple wizard-based protocol

Specifications

Measuring Principle

  • Raman spectroscopy

Ambient Temperature

  • Operating: -20 to 50 °C (-4 to 122 °F)
  • Recommended storage: 15 to 25 °C (59 to 77 °F)

Relative Humidity

  • Recommended storage: < 10 % humidity, non-condensing

Unit Dimensions (W x H x D)

  • 83 x 66 mm (3.27 x 2.60 inch)

Weight

  • 0.54 kg (1.20 lbs)

Spectral Intensity Reference

  • NIST SRM glass

Spectral Intensity Output Repeatability (at time of certification)

  • < ±2 %

Long term Spectral Uncertainty (at any wavelength)

  • ±2.85 %

Spectral Intensity Calibration Range

  • 534.5 to 694.0 nm
show more