Chromate Analyzer Liquiline System CA80CR
Brand: Endress+HauserChromate Analyzer Liquiline System CA80CR
Brand: Endress+Hauser
Liquiline System CA80CR offers highly accurate online measurement for optimized chromate removal in industrial wastewater. It supports you in achieving environmentally compliant electroplating and tanning processes. Like all Liquiline System analyzers, CA80CR enables plug & play of up to four Memosens sensors - minimizing the installation effort. Automatic calibration and cleaning plus advanced diagnostics with remote access ensure consistent process safety.
Liquiline System CA80CR supports you in improving the environmental performance of your processes by:
- Monitoring and optimizing the cleaning capacity of industrial wastewater treatment plants.
- Effluent monitoring in tanneries and electroplating plants.
- Monitoring of membrane filtration.
- Limit monitoring in drinking water applications.
- Standardized colorimetric measuring principle - diphenyle carbazide method - guarantees compliance with discharge regulations
- Reduced operating costs through automatic calibration and cleaning and low reagent consumption
- Simple maintenance: no tools required
- Advanced diagnostics with remote access for increased process safety
- Fast commissioning and plug & play thanks to Memosens technology and user-friendly operation
- Easy upgrade of functionality - even to a complete measuring station - simply by adding modules and connecting Memosens sensors. Reduces installation effort
Specifications
Brand
Measuring Principle
- Colorimetric
Characteristic
- Analyzer for Chromate (Chrome VI) in aqueous solutions
Size
- Housing (open version): 793 x 530 x 417 mm (31.22 x 20.87 x 16.42 in)
- Housing (closed version): 793 x 530 x 463 mm (31.22 x 20.87 x 18.23 in)
- Housing with base: 1723 x 530 x 463 mm (67.83 x 20.87 x 18.23 in)
Design
- Open design, cabinet and stand housing
- High-Performance plastic ASA-PC, additional stand coated steel
Process Temperature
- 4 to 40 °C (39 to 104 °F)
Ambient Temperature
- 5 to 40 °C (41 to 104 °F)
- Outdoor version: - 20 to 40 °C (-4 to 104 °F)
Process Pressure
- At atmospheric pressure, < 0.2 bar absolute
Sample Flow Rate
- Min. 5 ml/min (0.17 fl.oz/min)
Consistency of the Sample
- Suspended solids content
- Turbitity < 50 NTU, aqueous, homogenized
Specials
- Easy upgrade to measuring station with up to four digital Memosens sensors
- Automatic calibration and cleaning
- User-configurable measuring, cleaning and calibration intervals
- Optional cooling module for longer reagent shelf life
- 2 channel version optional
- User-definable measuring ranges
- Modular design for easily extensible functionality
- Digital communication for remote access
Application
- Limit value monitoring of Chromate in potable water and waste water
- Common applications are industrial waste water Monitoring for tanneries and the galvanic industry
Power Supply
- 100 to 120 VAC / 200 to 240 VAC ± 10%
- 24 VDC ± 10%
- 50 ± 1 or 60 ± 1.2 Hz
Output / Communication
- 2x 0/4 to 20 mA
- Modbus RS485/TCP (optional)
- Ethernet (optional)
- Alarmrelay
Input
- 1 or 2 measuring channels
- 1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)
Measuring Range
- 0,03 to 2.5 mg/l Cr(VI)
- 0.2 to 5.0 mg/l Cr(VI)
- 0.2 to 5.0 mg/l with dilution function to maximum 10 to 250 mg/l Cr(VI)
Consumables
- Reagents and standard solutions CY80CR as well as cleaner CY800 are necessary for the operation
- Regular maintenance is done with the parts of the maintenance kit CAV800