Raman Rxn2 Analyzer
Brand: Endress+HauserRaman Rxn2 Analyzer
Brand: Endress+Hauser
Adeptly harness the power of Raman spectroscopy with the Raman Rxn2 analyzer. Designed for use in analytical laboratories with model transfer capabilities, the Raman Rxn2 is heavily relied on for routine sample identification, support of R&D projects, early process development, and scale-up settings for in situ analysis. Available as a benchtop or on a mobile wheeled cart , the Raman Rxn2 offers location convenience and portability for process development laboratories.
- Chemical: polymerization, extrusion monitoring, polymer blending, catalysis investigation
- Pharmaceutical: reaction chemistry, yield crystallization, polymorph, blending, drying, granulations
- Biopharmaceutical: bioprocess monitoring and control, PAT/QbD applications, cell culture, fermentation, downstream operations, cGLP/cGMP compatible
- Food & Beverage: cultured meat, fermentation, oil or fat quality
- Reliable real-time, in situ measurements
- Intuitive, embedded control software via touchscreen or remote interface
- Convenience of a single base unit supporting up to four probes
- Sequential operation for fast analysis per channel and programmable channel interrogation
- Converts acquired Raman spectra into process knowledge using built-in multivariate predictors
- Suitable for outputs into hazardous area/classified environments
- Proven model transferability to process development and manufacturing scales
Specifications
Brand
Measuring Principle
- Raman spectroscopy
Laser Wavelength
- Starter: 785 nm
- Base Model: 532 nm, 785 nm, 1000 nm
- Hybrid: 785 nm
Spectral Coverage
- Starter 785 nm: 300-3300 cm-1
- Base model 785 nm: 100-3425 cm-1
- Base model 532 nm: 100-4350 cm-1
- Base model 1000 nm: 200-2400 cm-1
- Hybrid 785 nm: 150-1890 cm-1
Spectral Resolution
- Starter: 10 cm-1 (average)
- Base model (average): 5 cm-1 (532 nm), 4 cm-1 (785 nm), 6 cm-1 (1000 nm)
- Hybrid: 4 cm-1 (785 nm) average
Channels
- Starter: Single channel
- Base Model: Up to four channels
- Hybrid: Up to two channels
Ambient Temperature
- Operating: 15 to 30 °C (59 to 86 °F)
- Storage: -15 to 50 °C (5 to 122 °F)
Relative Humidity
- 20 to 80 %, non-condensing
Input Voltage
- 100-240 V, 50-60 Hz, ±10%
Power Consumption (W)
- 400 (max)
- 250 (typical start-up)
- 120 (typical running)
Warm Up Time (minutes)
- 120
Unit Dimensions (W x H x D)
- Benchtop model: 279 x 483 x 592 mm (10.99 x 19.02 x 23.31 inch)
- Cart model: 685 x 1022 (to tabletop) x 753 mm (26.97 x 40.24 (to tabletop) x 29.65 inch)
Weight
- Base model: 32 kg (71 lbs)
- Car model: 93 kg (205 lbs)
Sampling Probe Compatibility
- Starter: Raman Rxn-10 (with accessory optics)
- Base model: Raman Rxn-10 (with accessory optics), Rxn-40, Rxn-41, Rxn-45, Rxn-46
- Hybrid: Channel 1 - Raman Rxn-20 (with accessory optics)
- Hybrid: Channel 2 - Raman Rxn-10 (with accessory optics), Rxn-40, Rxn-41, Rxn-45, Rxn-46
Automation Interface
- OPC
- Modbus
- HTTPS
- (contact us for other options)
Installation Options
- Benchtop (standard) or mobile wheeled cart
Hazardous Area Certifications
- ATEX, CSA, IECEx