OXY5500 Oxygen Analyzer
Brand: Endress+HauserOXY5500 Oxygen Analyzer
Brand: Endress+Hauser
The OXY5500 oxygen analyzer is a compact, stand-alone, single channel analyzer that uses quenched fluorescence (QF) technology to reliably measure oxygen in natural gas and gas processing applications. A sensor probe is inserted into the process stream and is connected to the controller by an optical fiber. Featuring an LCD display and built-in data logger, the OXY5500 also has full sample conditioning systems available as well as software for a PC interface.
- Natural gas quality measurements in production, storage, transportation, and distribution
- Monitoring in natural gas, upstream of production and gathering sites
- Measurable range from ppm levels to 0-20% O2
- Trace measurements as low as 0-10 ppmv oxygen in hydrogen for electrolyzer applications
- Ideal for CCUS applications since optical oxygen measurement is unaffected by CO2or H2S in the process
- Easy to navigate display and menu
- Small optical sensor with no membrane or consumable chemicals
- Optical measurement with fast, continuous response
- Excellent long-term stability
- Not affected by H2S
- No moving parts, simple operation
- Data logging for 30 days
Specifications
Brand
Measuring Principle
- Quenched Fluorescence
Channels
- 1
Analyte and Measurement Ranges
- O2 (Oxygen):
- 0 to 100 ppmv (minimum)
- 0 to 20% volume (maximum)
Measured Variables
- Concentration
- Cell Pressure
- Cell Temperature
Ambient Temperature
- -20 to 50°C (-4 to 122°F)
Operating Pressure Range
- Inlet Pressure: 140 to 275 kPa (20 to 40 psig)
- Sample Cell: 800 to 1200 mbara
Analyzer Wetted Materials
- 304 & 316 stainless steel
- FKM O-Rings
Power Supply
- 108 to 253 VAC, 50/60 Hz; 5.3 Watts at 120VAC; 6.6 Watts at 240VAC
- 9 to 30 VDC (CSA), 18 to 30 VDC (IECEx/ATEX); 4.7 Watts at 24VDC
Communication
- Analog Output: Two 4-20 mA outputs
- Analog Input: One 4-20 mA input (sample pressure)
- Fieldbus: RS-232C, RS-485, & Ethernet 10/100 with Modbus
- Output Relays: Qty 2, 250 mA max load (Concentration and Fault Alarms)
- USB 2.0 Works With Service Software Only
- 4 GB Internal Memory with Internal Data Logging
Housing Materials
- Electronics: 304 or 316L stainless steel
Hazardous Area Approvals
- ATEX / IECEx - Zone 2
- CSA Class I, Division 2
Degree of Protection
- IP64, Type 4X
Product Safety
- CE
- RCM
- FCC