Water Hardness Analyzer Liquiline System CA80HA
Brand: Endress+HauserWater Hardness Analyzer Liquiline System CA80HA
Brand: Endress+Hauser
Liquiline System CA80HA offers precise online analysis of water hardness in drinking water and process water. It helps you optimize the control of water softening processes such as ion exchange or reverse osmosis and ensures optimal quality of products that are influenced by water hardness. The analyzer's automatic calibration and cleaning and its low reagent consumption reduce operating costs while advanced diagnostics with remote access ensure consistent process safety.
- To optimize reverse osmosis and ion exchanger systems
- To classify water hardness in drinking water
- To ensure process water quality in production plants
- The water hardness analyzer uses the phthalein purple method, ensuring direct comparability to lab results
- Fast commissioning with plug & play thanks to Memosens technology and user-friendly operation
- Reduces operating costs thanks to automatic calibration and cleaning, low reagent consumption and long reagent lifetime with optional cooling module
- Easy upgrade of functionality to a complete measuring station - by adding modules and connecting Memosens sensors.
- Advanced diagnostics with remote access for increased process safety
- Easy, tool-free maintenance
Specifications
Brand
Measuring principle
- Colorimetric
Characteristic
- Process analyzer for hardness in aqueous solutions
- Comply with standard colorimetric measuring principle - phthalein purple method
Size
- Housing (open version): 793 x 530 x 417 mm / 31.22 x 20.87 x 16.42 in
- Housing (closed version): 793 x 530 x 463 mm / 31.22 x 20.87 x 18.23 in
Design
- Housing with base: 1723 x 530 x 463 mm / 67.83 x 20.87 x 18.23 in
- Open design, cabinet and stand housing
- High-Performance plastic ASA-PC, additional stand coated steel
Process temperature
- 4 to 40 °C (39 to 104 °F)
Ambient temperature
- 5 to 40 °C (41 to 104 °F)
Process pressure
- Unpressurized
Sample flow rate
- Min. 5 ml/min (0.17 fl.oz/min)
Consistency of the sample
- Low solid content (< 50NTU), aqueous homogeneous
Application
- Total hardness monitoring in reverse osmosis systems inlets, after ion exchange systems
- Monitoring of water preparation in processes and drinking water
Power supply
- 100 ... 120 VAC / 200 ... 240 VAC ± 10%
- 24 VDC ± 10%
- 50 ± 1 oder 60 ± 1,2 Hz
Output / communication
- 2x 0/4 ... 20 mA
- Webserver, Modbus, Ethernet/IP, Profibus DP
Input
- 1 or 2 measuring channel
- 1 ... 4 digital sensor inputs for sensors with Memosens protocol (optional)
Measuring range
- 0 to 80 mg/l CaCO3
- 0 to 80 mg/l with dilution function to maximum 16 to 1600 mg/l CaCO3
Consumables
- Reagents and standard solutions CY80HA as well as cleaner CY800 are necessary for the operation
- Regular maintenance is done with the parts of the maintenance kit CAV800