Proline T-Mass A 150 Thermal Mass Flowmeter

Brand: Endress+Hauser

Proline T-Mass A 150 Thermal Mass Flowmeter

Brand: Endress+Hauser
Lean Lean

The t-mass A 150 inline device is specially designed for the cost-effective measurement of utility gases, in particular compressed air. It is a trending device aimed at sub-metering applications. its 4-wire technology is contained within a rugged compact aluminum housing. It is orderable either with display or as a blind version. Customer-specific settings are saved on the display and can be transferred from one device to another by means of the display.

  • Measuring principle is characterized by a high operable flow range and direct mass flow measurement
  • Suitable for air, nitrogen, carbon dioxide and argon in small line sizes

Device properties

  • Inline version: nominal diameter DN 15 to 50 (1/2 to 2")
  • Process pressure up to PN 40, Class 300
  • A variety of process connections available
  • Device in compact version with DC 24 V power supply
  • 4-20 mA HART, pulse/frequency/switch output
  • Compact and robust transmitter
Features
  • Suitable for air, nitrogen, carbon dioxide and argon in small line sizes
  • Optimal process monitoring – easy measurement even at low pressures and flow velocities
  • Cost-effective measurement – easy installation, negligible pressure loss and maintenance-free
  • Reliable flow trending – multivariable measurement
  • Fast and efficient commissioning – guided operating menus
  • High plant availability – self-diagnostics and error monitoring
  • Automatic recovery of data for servicing

Specifications

Measuring Principle

  • Thermal

Product Headline

  • The flowmeter for cost-effective measurement and easy monitoring of utility gases
  • Suitable for air, nitrogen, carbon dioxide and argon in small line sizes

Sensor Features

  • Optimal process monitoring – easy measurement even at low pressures and flow velocities. Cost-effective measurement – easy installation, negligible pressure loss and maintenance-free. Reliable flow trending – multivariable measurement
  • Inline version: DN 15 to 50 (1/2 to 2"). Process pressure up to PN 40, Class 300. A variety of process connections available

Transmitter Features

  • Fast and efficient commissioning – guided operating menus. High plant availability – self‐diagnostics and error monitoring. Automatic recovery of data for servicing
  • Device in compact version with DC 24 V power supply. 4-20 mA HART, pulse/frequency/switch output. Compact and robust transmitter

Nominal Diameter Range

  • DN 15 to 50 (1/2 to 2")

Wetted Materials

  • Transducer: 1.4404 (316L)
  • Insertion tube: 1.4404 (316L)
  • Measuring tube: 1.4404 (316L); 1.4435 (316L)
  • Connection: 1.4404 (F316/F316L); 1.4404 (316L); 1.4435 (316L)

Measured Variables

  • Mass flow, temperature, corrected volume flow, FAD volume flow

Max. Measurement Error

  • 3 % o.r.
  • 4 % o.r.
  • 5 % o.f.s.
  • depending on chosen option of ordering feature "Calibration flow"

Measuring Range

  • 0.5 to 910 kg/h (1.1 to 2002 lb/h)
  • 0.5 to 1365 kg/h (1.1 to 3003 lb/h)
  • for air, depending on chosen option of ordering feature "Calibration flow"

Max. Process Pressure

  • PN 40, Class 300

Medium Temperature Range

  • –40 to +100 °C (–40 to +212 °F)

Ambient Temperature

  • –40 to +60 °C (–40 to +140 °F)

Transmitter Housing Material

  • AlSi10Mg, coated

Degree of Protection

  • IP66/67, type 4X enclosure

Display/Operation

  • 4‐line display with push Buttons
  • Configuration via local display and operating tools possible

Outputs

  • 4‐20 mA HART (active)
  • Pulse/frequency/switch output (passive)

Inputs

  • Status input

Digital Communication

  • HART

Power Supply

  • DC 18 to 30 V

Hazardous Area Approvals

  • ATEX, IECEx, cCSAus

Metrological Approvals and Certificates

  • Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR

Pressure Approvals and Certificates

  • PED, CRN
show more