iTHERM ModuLine TM411 Hygienic Modular Thermometer

Brand: Endress+Hauser

iTHERM ModuLine TM411 Hygienic Modular Thermometer

Brand: Endress+Hauser
Extended Extended

The modular sanitary thermometer iTHERM TM411 with digital display delivers highly accurate, reliable and fast temperature measurements in process lines, vessels, tanks, fermenters and more. The high-performance device is designed for the strictly regulated food and life sciences industries. Configure your hygienic instruments from a selection of thin film RTDs with QuickSens or StrongSens technology or 2x Pt100 wire wound temperature sensors, smart transmitters and hygienic process connections.

  • Designed for use in hygienic and aseptic applications in food, beverage and life sciences
  • Measuring range: –200 to 600 °C (-28 to 1 112 °F)
  • Pressure range up to 40 bar (580 psi)
  • Protection class: up to IP69K
  • Remote parametrization with secureBluetooth® temperature transmitters
  • International approvals: Certified for use in hazardous areas, explosion protection as per ATEX/IECEx, CSA, FM, EAC Ex, JPN Ex, NEPSI; Hygiene standards according to 3-A, EHEDG, ASME BPE, FDA, TSE Certificate of Suitability
Features
  • Optimum process control: Fast sensor response times with iTHERM QuickSens technology (t90: 1.5 s)
  • Up to 75% faster and safer recalibrations with iTHERM QuickNeck – tool-free quick-release for instant insert removal
  • High sensor availability under rough conditions: Vibration resistant RTD technology iTHERM StrongSens (> 60g)
  • Easy product selection, configuration and maintenance
  • Increase productivity and product safety in sanitary applications with highly accurate and reliable measurements
  • Thermowells, hygienic process connections and housings expertly manufactured from high-grade materials treated for low surface roughness

Specifications

Measuring Principle

  • Resistance Temperature Detector

Characteristic / Application

  • metric style
  • modular temperature assembly
  • hygienic/aseptic design (3-A®, EHEDG, ASME BPE, FDA)
  • hygienic process connections
  • with neck
  • suitable for hazardous areas
  • incl. protection tube
  • iTHERM QuickSens for fastest response time
  • iTHERM StrongSens for most robust design
  • iTHERM QuickNeck for easy and cost saving recalibration

Thermowell / Protection Tube

  • without
  • incl. thermowell
  • incl. Tee-/elbow thermowell

Insert / Probe

  • mineral insulated (MI), flexible
  • pipe version, isolated wires, not flexible

Outer Diameter Protection Tube / Insert

  • 6.0 mm (0.24")
  • 9.0 mm (0.35")
  • 12.7 mm (0.5")

Material of Protection Tube/ Thermowell

  • 1.4435 delta-ferrit < 1%
  • 316L

Process Connection

  • compression fitting
  • weld in adapter
  • clamp connections acc. to ISO2852
  • screwed pipe joint acc. to DIN 11851
  • aseptic screwed pipe joint acc. to DIN 11864-1
  • metallic sealing system
  • thread acc. to ISO 228 for liquiphant adaptor
  • APV Inline
  • Varivent
  • Ingold
  • SMS 1147
  • Neumo Biocontrol
  • T- and elbow thermowell

Tip Shape

  • straight
  • reduced
  • tapered

Surface Roughness Ra

  • 0.76 μm (29.92 μin)
  • 0.38 μm (14.96 μin)
  • 0.38 μm (14.96 μin) electropolished

Operating Temperature

  • PT100: -200 °C … 600 °C (-328 °F … 1,112 °F)
  • iTHERM StrongSens: -50 °C … 500 °C (-58 °F … 932 °F)
  • iTHERM QuickSens: -50 °C … 200 °C (-58 °F … 392 °F)

Max. Process Pressure (static)

  • at 20 °C: 40 bar (580 psi)

Accuracy

  • class A acc. to IEC 60751
  • class AA acc. to IEC 60751

Response Time

  • depending on configuration
  • iTHERM QuickSens: t90 = 1.5 s
  • iTHERM StrongSens: t90 = 9.5 s

Integration Head Transmitter

  • yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)

Ex - Approvals

  • ATEX Ex ia, Ex nA
  • ATEX IECEx Ex ia, Ex ta/tb
  • CSA C/US IS, NI
  • FM IS, NI
  • NEPSI Ex ia
  • JPN Ex ia
  • EAC Ex ia
  • UKCA Ex ia, Ex nA, Ex tc

Certification

  • SIL (transmitter only)
show more