Raman Rxn-10 Probe

Brand: Endress+Hauser

Raman Rxn-10 Probe

Brand: Endress+Hauser

Local Support: Authorized Distributor Serving Alabama, Mississippi, Louisiana, Florida, New York, Connecticut, Massachusetts, Vermont, New Hampshire, Maine, Rhode Island, Puerto Rico, US Virgin Islands and parts of Tennessee and Arkansas.

Extended Extended

Designed for product and process development, the Rxn-10 probe is the workhorse of the Raman probe portfolio. It is trusted to deliver high performance Raman measurements across a wide spectral range. It is also compact, lightweight, and flexible, offering multi-purpose convenience for solids and liquids analysis in the laboratory environment. The Rxn-10 probe accepts a variety of interchangeable optics, making it a highly versatile and easily adaptable instrument in your laboratory toolbox.

  • Chemical: reaction monitoring, blending, catalysis, hydrocarbon speciation, process unit optimization
  • Polymer: polymerization reaction monitoring, extrusion monitoring, polymer blending
  • Pharmaceutical: API reaction monitoring, crystallization
  • Biopharmaceutical: monitoring, optimization, and control of cell culture, fermentation, and downstream applications
  • Food & Beverage: zonal heterogeneity mapping of meats and fish
Features
  • Multipurpose use for both solids and liquids measurement
  • Lightweight and compact
  • Integrated laser safety interlock, including “laser on” indication and probe shutter
  • Flexible output compatible with a range of sampling options
  • Easy switching of non-contact, immersion, and bioprocessing optics to suit a variety of applications
  • Wide spectral range, including access to the critical low-wavenumber region

Specifications

Product Family

  • Raman

Measuring Principle

  • Raman Spectroscopy

Laser Wavelength

  • With immersion optic: 532 nm, 785 nm, 1000 nm
  • With bIO-Optic or Raman optic system for single use: 785 nm, 1000 nm
  • With bio multi optic and bio sleeve or Raman flow assembly: 785 nm

Spectral Coverage

  • Probe spectral coverage is limited by the coverage of the analyzer being used

Ambient Temperature

  • Probe: –10 to 70 °C (14 to 158 °F)

Relative Humidity

  • 20 to 60 %, non-condensing

Maximum Laser Power Into Probehead (mW)

  • < 499

Sample Interface

  • Based on the sampling optic selected

Body & Window Materials

  • Rxn-10 probe body: 6061 aluminum, 316L stainless steel, and 303 stainless steel

Fiber Optic Cable

  • Design: PVC jacketed, proprietary construction
  • Connections: proprietary electro-optic (EO), or FC to EO fiber converter(s) for non-embedded systems
  • Temperature: -40 to 70 °C (-40 to 158 °F)
  • Standard length: 5000 mm (196.86 inch); 25000 mm (984.26 inch)
  • Also available in 5000 mm (196.86 inch) increments
  • Minimum bend radius: 152.4 mm (6 inch)

Length

  • Rxn-10 probe (including fiber cable bend radius): 356 mm (14.02 inch)

Diameter

  • Rxn-10 probe: 19 mm (0.75 inch)

Working Distance

  • Based on the sampling optic selected
show more