TMR35 Compact Thermometer
Brand: Endress+HauserTMR35 Compact Thermometer
Brand: Endress+Hauser
With cost efficiency, optimal use of space, reliable operation, easy commissioning and hygienic design the TMR35 fulfills all the requirements in the Food & Beverage as well as in the Life Science industries. A variety of hygienic process connections leads to a high flexibility.
The TMR35 compact thermometers are used to measure temperatures from -50 °C to +150 °C (-58 °F to +302 °F), or up to +200 °C (+392 °F) with neck. The most common installation locations are tanks and pipes. TMR35 with process connections for hygienic applications.
Features- Small, compact design made entirely of stainless steel
- Extremely short response times
- Highly accurate even with short insertion lengths
- 4-wire, Pt100 or PC-programmable transmitter with 4 to 20 mA output
- Configuration and visualization with ReadWin®2000 PC operating software, which is free of charge
- Breakdown information in event of sensor break or sensor short-circuit, adjustable as per NAMUR NE43
- Hygienic compatible design with 3-A marking and EHEDG certification;Marine approval
Specifications
Brand
Measuring Principle
- Resistance Temperature Detector
Characteristic / Application
- metric style
- compact temperature probe
- fast response time
- hygienic design
- hygienic process connection
- with or without neck
Thermowell / Protection Tube
- without (not intended to be used with thermowell)
- to be used with thermowell
- incl. Tee-/elbow thermowell
Insert / Probe
- pipe version, isolated wires, not flexible
Outer Diameter Protection Tube / Insert
- 3.0 mm (0.12'')
- 6.0 mm (0.24'')
Max. Immersion Length on Request
- up to 600.00 mm (23.62'')
Material Protection Tube/ Thermowell
- 1.4404 (316L)
Process Connection
- hygienic:
- conical sealing system G1/2" (metal)
- clamp connections acc. to ISO 2852 / DIN 32676
- screwed pipe joint acc. to DIN 11851
- Varivent
Tip Shape
- straight
Surface Roughness Ra
- 0.4 μm (16 μin.)
- 0.8 μm (32 μin.)
Operating Temperature
- PT 100: -50 °C ...200 °C (-58 °F ...392 °F)
Max. Process Pressure (static)
- at 20 °C: 100 bar (1,450 psi)
Accuracy
- class A acc. to IEC 60751
Response Time
- t50 = 1 s
- t90 = 2 s
Integration Head Transmitter
- no (4…20mA Signal)
Certification
- 3A
- Marine approval
- UL 3111-1