Raman bIO-Optics

Brand: Endress+Hauser

Raman bIO-Optics

Brand: Endress+Hauser
Extended Extended

Raman bIO-Optics are versatile immersion optics used in conjunction with the Rxn-10 probehead and compatible with standard bioreactor ports. The fixed focus design provides long-term measurement stability along with superior signal performance, essential for transferable, high performance Raman-based bioprocess analysis. Available in industry standard lengths, Raman bIO-Optics are ideally suited for both glass and plastic benchtop bioreactor/fermenter applications requiring headplate entry.

  • Cell culture:glucose, lactate, glutamine, amino acids, cell density, titer
  • Fermentation:glycerol, methanol, ethanol, sorbitol, biomass
  • Downstream:aggregation, protein crystallization, formula stability, product CQA, protein concentration, buffer excipients
Features
  • Immersion optics for multi-component bioprocess measurement
  • Compatible with standard bioreactor ports
  • Fixed-focus design
  • Available in 12 mm diameter with Pg 13.5 threaded connector
  • 120, 220, 320, or 420 mm lengths
  • Autoclavable

Specifications

Measuring Principle

  • Raman spectroscopy

Laser Wavelength

  • 785 nm, 1000 nm

Sampling Probe Compatibility

  • Raman Rxn-10 probe

Sample Interface

  • Temperature: –30 to 150 °C (-22 to 302 °F)

Pressure

  • Max pressure: 13.8 barg (200 psig)

Wetted Materials

  • Body: 316L stainless steel
  • Window: Proprietary material, optimized for bioprocesses
  • Process connection: PG13.5
  • Surface finish: Ra 15 with electropolish
  • Adhesive: USP Class VI and ISO 10993 compatible

Length

  • 120, 220, 320, or 420 mm
  • 4.73, 8.67, 12.60, 16.54 inch

Diameter

  • 12 mm (0.48 inch)

Sterilization Method

  • Autoclave

Process Connection

  • PG13.5
show more