iTHERM CableLine TH12 Cable Probe

Brand: Endress+Hauser

iTHERM CableLine TH12 Cable Probe

Brand: Endress+Hauser

The Pt100 cable probe is easy to install and provide a high operational safety due to reliable and accurate temperature measurement in common processes. Without additional thermowell the probe sheath is directly in contact with the process medium. This enables the cable probe to detect rapid temperature changes fast and accurate.

The TH12 temperature sensors are RTD's designed for use in the process industry or factory automation type of environment. They are made up of a measurement probe with a RTD, insulation and sheath.

Among other applications the sensors can be used on:

  • Heat exchangers
  • Power & recovery areas
  • Carbon regeneration furnaces
  • Dryers
  • Flue gas
  • Compressor stations
  • Process reactors
Features
  • High flexibility due to customized immersion length
  • Fast response time with reduced/tapered tip form
  • Simplified model structure: Competitively priced, offers great value. Easy to order and reorder
  • Improved galvanic isolation on most devices (2 kV)

Specifications

Measuring Principle

  • Resistance Temperature Detector

Characteristic / Application

  • US style
  • cable probe
  • process connection as compression fitting
  • without neck

Thermowell / Protection Tube

  • without (not intended to be used with thermowell)

Insert / Probe

  • mineral insulated (MI), flexible
  • PTFE-insulated, rigid

Outer Diameter Protection Tube / Insert

  • 1/8'' (3.18 mm)
  • 3/16'' (4.76 mm)
  • 1/4'' (6.35 mm)
  • 3/8'' (9.53 mm)

Max. Immersion Length on Request

  • up to 96" (2439 mm)
  • others on request

Material Protection Tube/ Thermowell

  • Sensor sheath
  • 316/316L

Optional Coating

  • Not applicable

Process Connection

  • Compression fitting: NPT1/8", NPT1/4"

Tip Shape

  • straight
  • reduced

Surface Roughness Ra

  • Not defined

Operating Temperature

  • PT100 WW: -200 °C … 600 °C (-328 °F … 1.112 °F)
  • PT100 TF: -50 °C … 200 °C (-58 °F … 392 °F)

Max. Process Pressure (static)

  • at 20 °C: 40 bar (580 psi)
  • depends on configuration

Accuracy

  • class A acc. to IEC 60751
  • class B acc. to IEC 60751

Response Time

  • 63% rt = 2.0 s

Integration Head Transmitter

  • Not applicable
show more